Description
Product Description of Bionime GM700 Glucometer
|
Test strip |
GS700 |
|
Coding |
auto-coding |
|
Enzyme / measurement technology |
GDH / electrochemical sensor |
|
Blood sample source |
capillary, venous, arterial and neonatal whole blood |
|
Minimum sample volume |
0.75 µL |
|
HCT range |
glucose concentrations ≦ 200 mg/dL, HCT range 20 – 70% Glucose concentrations > 200 mg/dL, HCT range 20 – 60% |
|
Measuring range |
0.6 – 33.3 mmol/L (10 – 600 mg/dL) |
|
Meter dimensions (L x W x H) |
96.0 x 46.0 x 17.5 mm (3.78 x 1.81 x 0.68 inch) |
|
Display area (L x W) |
52.6 x 32.0 mm (2.07 x 1.25 inch) |
|
Weight (with battery) |
57.0 ± 5 g |
|
Power supply |
1 × CR2032 battery |
|
Memory capacity |
1,000 test results with date and time |
|
Operating temperature |
6 – 44℃ (43 – 111°F) |
|
Operating relative humidity |
10 – 90% |
|
Meter storage conditions |
-10 – 60°C (14 – 140°F) |
|
Test strip storage conditions |
At 4 – 30°C (39 – 86°F), and below 90% relative humidity |
|
Accuracy |
for glucose concentrations < 5.5 mmol/L (100 mg/dL), 95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements. For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), 95% of measured glucose values within ± 15% of reference measurements. |
|
Precision |
for glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL) For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5% |
- 01 Year warranty














Reviews
There are no reviews yet.